Cardiac Marker Control Calibration Verification Set, List number 136604/06F15-02. Used to verify the accuracy...
Abbott Point Of Care Inc. began a recall of Cardiac Marker Control Calibration Verification Set, List number 136604/06F15-02. Used to verify the accuracy of results over the measurement range of the... on . The reason FDA records is "Presence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1864-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Cardiac Marker Control Calibration Verification Set, List number 136604/06F15-02. Used to verify the accuracy of results over the measurement range of the i-STAT cTnI test.
- Product code
- JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
- Reason for recall
- Presence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Recall Notice letters with business reply cards were sent Fed ex on December 17, 2010 to all accounts. International notifications were sent on December 21, 2010. The letter identified the affected product and lot numbers along with a background explanation of the issue. The firm recommends that customers remove the cap from the vial; squeeze the vial 10 times, prior to thawing; replace cap on the vial; and then follow the remaining steps for use as outlines in section 14-12 of the i-STAT 1 System Manual. Customers are to complete and return the Business Reply card to acknowledge receipt of the letter. If the affected product had been further distributed then a copy of the letter should be provided to those customers as well. If additional information is needed, customers should contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or their Abbott Point of Care representative.
- Quantity in commerce
- 557 units total
- Distribution
- Worldwide Distribution
- Product codes and lots
- Cardiac Marker Control Calibration Verification Set - Lot number M10208