Z-1670-2011
Ovation Hip Stem, REF 201-0002, Hip Stem Inserter, Non-Sterile, Ortho Development, Draper, Utah 84020. The...
Ortho Development Corp. / initiated 2011-01-11 / Terminated / root cause: manufacturing process
Ortho Development Corporation began a recall of Ovation Hip Stem, REF 201-0002, Hip Stem Inserter, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip... on . The reason FDA records is "Component of total hip arthroplasty may break during surgical procedure". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1670-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho Development Corp.
- Product
- Ovation Hip Stem, REF 201-0002, Hip Stem Inserter, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. The hip stem inserter is an instrument used to seat the hip stem in the femur. There is a notch in the top of the hip stem, and the hip stem inserter has a corresponding tip at the end of the instrument. The tip of the inserter is designed to fit into the notch of the hip stem, as a guide. Once this is inserted, the surgeon will tap on the opposite end of the inserter with a hammer to seat the press-fitted stem assembly.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- Component of total hip arthroplasty may break during surgical procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Ortho Development Corporation notified Field representatives of the issue and conditions that could lead to failure by e-mail on December 14, 2010. Redesigned product is being produced for replacement. For questions regarding this recall call 801-619-3450.
- Quantity in commerce
- 33 units
- Distribution
- Nationwide Distribution including AZ, CA, CO, FL, MI, NV, OR, and UT.
- Product codes and lots
- All lots