Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is...
Computerized Medical Systems Inc began a recall of Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is used to create treatment plans for any cancer... on . The reason FDA records is "Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2561-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco Radiation Treatment Planning Workstation. Monaco Release 2.03.00 and above. The Monaco RTP system is used to create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Monaco: For VMAT plans, the measured dose of radiation delivered is lower than the software calculated dose by about 5% in Monaco Release 2.03.00 and 2.04.00.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Computerized Medical Systems, Inc. sent a "USER NOTICE" on January 13, 2011, to all affected users. This notice identified the product, the problem, and the action needed to be taken by the customer. A return postcard is included for the customer to confirm receipt of the user notification. The firm intends for correction in the dose calulation engine to appear in the Monaco Release 3.00 in March 2011. Any further questions please call (403) 830-8023.
- Quantity in commerce
- 94
- Distribution
- Worldwide Distribution-- USA (nationwide) and countries of Austria, Australia, Belarus, Canada, China, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Mexico, Netherlands, Poland, Singapore, Slovenia, South Korea, Spain, Switzerland, Taiwan, Thailand and UK.
- Product codes and lots
- Release 2.03.00 and above.