Z-1043-2011

GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM ( SENO ESSENTIAL) The expected usage of this product is...

GE Healthcare / initiated 2010-07-20 / Terminated

GE Healthcare, LLC began a recall of GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM ( SENO ESSENTIAL) The expected usage of this product is for the same clinical applications as traditional... on . The reason FDA records is "It was noticed that 3 units received in Japan were missing the rating plate on the X-ray tube cover. There is a potential that other X-ray tube covers are missing the rating plate". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1043-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE SENOGRAPHE ESSENTIAL FULL FIELD MAMMOGRAPHY SYSTEM ( SENO ESSENTIAL) The expected usage of this product is for the same clinical applications as traditional mammographic film/screen systems. It generates mammographic images which can be used for screening and diagnosis of breast cancer.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
It was noticed that 3 units received in Japan were missing the rating plate on the X-ray tube cover. There is a potential that other X-ray tube covers are missing the rating plate.
FDA root cause
Radiation Control for Health and Safety Act
Action
An Important Electronic Product Radiation Warning letter was issued to customers identifying the affected product and describing the defect along with the related hazards. The letter states that there are no actions that are required to be taken on the part of the customer. A GE Healthcare Service Representative will apply the missing rating as necessary and free of charge. Questions or concerns regarding the letter should be directed to 1-800-437-1171 in the US.
Quantity in commerce
112 total
Distribution
Worldwide Distribution
Product codes and lots
All associated serial numbers.