Universal Insertion Tray Bundle, CV13530, Sterile, Centurion Medical Products Corp., Williamston, MI. These...
Centurion Medical Products Corporation began a recall of Universal Insertion Tray Bundle, CV13530, Sterile, Centurion Medical Products Corp., Williamston, MI. These kits are used for dressing changes on . The reason FDA records is "The firm is conducting a subrecall of kits containing Triad Groups alcohol prep pads. Triad initiated a recall of prep pads due to a concern about potential contamination of the products with an objectional orgainism...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1460-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Universal Insertion Tray Bundle, CV13530, Sterile, Centurion Medical Products Corp., Williamston, MI. These kits are used for dressing changes.
- Product code
- LRS I.V. Start Kit
- Reason for recall
- The firm is conducting a subrecall of kits containing Triad Groups alcohol prep pads. Triad initiated a recall of prep pads due to a concern about potential contamination of the products with an objectional orgainism, Bacillus cereus.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Centurion Medical Products, sent an "URGENT PRODUCT RECALL NOTICE" dated January 12, 2011 to affected customers via Certified Mail, Return Receipt Requested. The notice described the product, problem and action to be taken. The customers were instructed to identify and remove all inventory of the recalled kits in their possession; forward a copy of this notice to any customers that this product was further distributed to, and complete and return the attached Accountability Record form via fax to 517-546-3356 or email to: [email protected]. The notice states that a Centurion representative will contact them regarding retrieval of the product from the facility. Should your require additional information, please contact the Director of Quality Assurance & Regulatory Affairs at (517) 546-5400 ext. 1135.
- Quantity in commerce
- 68056 all kits
- Distribution
- Nationwide distribution: USA including state of: CT, ID, NM, MA, MO, MT, VA and SC.
- Product codes and lots
- 2010102550