Z-1401-2011

MAGNETOM Verio Nuclear magnetic resonance imaging

Siemens Medical Solutions USA, Inc. / initiated 2011-01-10 / Terminated

Siemens Medical Solutions USA, Inc began a recall of MAGNETOM Verio Nuclear magnetic resonance imaging on . The reason FDA records is "On some systems the gradient cable connections did not meet firm's specifications after installation". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1401-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MAGNETOM Verio Nuclear magnetic resonance imaging
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
On some systems the gradient cable connections did not meet firm's specifications after installation
FDA root cause
Other
Action
Siemens Medical Solutions USA, Inc sent customers a letter on January 10, 2011 notifying them of an Update Instruction and informing them that recalling firm would preventatively check the gradient cable connections of all systems. For questions regarding this recall call 610-219-6300.
Quantity in commerce
92
Distribution
Nationwide Distribution including AK, AZ, CA, CT, FL, HI, IL, IN, KY, MA, MD, MI, MN, MO, MT, NC, NJ, NM, NV, OH, OK, OR, PA, RI, TN, TX, VA, WI, and WV
Product codes and lots
Model number 10276755