Philips MultiDiagnost Eleva system, a diagnostic-imaging device intended for radiographic, fluoroscopic...
Philips Medical Systems North America Co. Phillips began a recall of Philips MultiDiagnost Eleva system, a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications. FCO... on . The reason FDA records is "A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1656-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips MultiDiagnost Eleva system, a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications. FCO 70800114.
- Product code
- IZI System, X-Ray, Angiographic
- Reason for recall
- A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 10/15/2010, Philips began sending out the URGENT-Device Correction Notice Failure Causing the Generator to Stop Working notification letter to their consignees. The consignees were informed that a short circuit may occur due to a faulty wire connection in the Supervisor Box Connector in the Generator, and it will cause the generator stops working. When the system fails during a critical interventional case, this may cause risk to patients (no Fluoro available). The systems involved in this recall are Allura Xper with the Velara Generator, Allura Integris with OMCP Generator, MultiDiagnost Eleva, and OmniDiagnost Eleva Imaging system. Consignees can contact Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 70800114, 70800115, 72200184, and 72200185.
- Quantity in commerce
- 89 units distributed in US
- Distribution
- Devices were distributed through out the US.
- Product codes and lots
- 505345, 505346, 506236, 530589, 530777, 532140, 532151, 532213, 532703, 533473, 533559, 534247, 535714, 535767, 536019, 536091, 536146, 536720, 536935, 537083, 537669, 537787, 539598, 539990, 540252, 541263, 543272, 544059, 545014, 545156, 545198, 545841, 546031, 546578, 549417, 551519, 551714, 553287, 553435, 553838, 554283, 554284, 554633, 554640, 556197, 556369, 557258, 41444651, 41445097, 41445127, 41445131, 41445221, 41445237, 41445464, 41445652, 41445800, 41445927, 41455927, 41643223, 41783749, 41849575, 41954854, 42549980, 42799914, 42892004, 42892024, 42892425, 43319037, 43386703, 43443091, 43453378, 43721197, 44146025, 44401636, 44653739, 44654723, 45629333, 47130091, 47190569, 47413010, 47607148, 47783855, 49136896, 49144094, 49903432, 50003342, 50037841, 50416595, and 51599597.