Haemonetics cardioPAT Cardiovascular - autotransfusion System Catalog Number: 02050-US
Haemonetics Corporation began a recall of Haemonetics cardioPAT Cardiovascular - autotransfusion System Catalog Number: 02050-US on . The reason FDA records is "Update Instructions for Use for the CardioPAT® Manual for Leak Detection. User may mis-interpret the air-leak rate displayed by the cardioPAT device when making a determination about chest-drain removal, which could...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1391-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Haemonetics Corporation
- Product
- Haemonetics cardioPAT Cardiovascular - autotransfusion System Catalog Number: 02050-US
- Product code
- CAC Apparatus, Autotransfusion
- Reason for recall
- Update Instructions for Use for the CardioPAT® Manual for Leak Detection. User may mis-interpret the air-leak rate displayed by the cardioPAT device when making a determination about chest-drain removal, which could result in the patient experiencing a pneumothorax.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Haemonetics issued a Field Correction Notice on 1/25/11 . with the Air Leak Detection addendum and index be inserted into the appropriate document, applied to all cardioPAT@ devices on site. The serial number for each cardioPAT device on site should be recorded on the attached acknowledgement and the completed acknowledgement should be returned to your Haemonetics Sales Consultant. A Haemonetics Sales Consultant will train on the content of these items and insert them into the Operator Manuals
- Quantity in commerce
- 309 units
- Distribution
- Worldwide distribution, including USA, Australia, Belgium, Canada, Switzerland, China, Germany, Finland, France, United Kingdom, Hong Kong, Israel, Italy, Japan, Netherlands, Russian Federation, and Sweden.
- Product codes and lots
- All serial numbers