Z-3091-2011

Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer...

Abbott Medical / initiated 2009-06-02 / Terminated / root cause: software

Advanced Neuromodulation Systems began a recall of Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based... on . The reason FDA records is "Product programmed with unapproved software was distributed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-3091-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting implantable pulse generators of the Eon, Genesis, and Renew family.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Product programmed with unapproved software was distributed.
Root cause tag
software
FDA root cause
Software change control
Action
St. Jude Medical Neuromodulation contacted all affected customers by telephone between June 2, 2009 and June 5, 2009. Customers were given information about the product, problem and action to be taken. All affected devices were returned to the firm as instructed.
Quantity in commerce
14 units
Distribution
Distributed in New Hampshire and Puerto Rico.
Product codes and lots
Serial numbers: 101565, 101544, 101555, 101548, 101561, 101551, 101563, 101546, 101556, 101553, 101554, 101558, 101545, and 101566