Z-3091-2011
Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer...
Abbott Medical / initiated 2009-06-02 / Terminated / root cause: software
Advanced Neuromodulation Systems began a recall of Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based... on . The reason FDA records is "Product programmed with unapproved software was distributed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-3091-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Software version 3.3 for Rapid Programmer devices, Model 3832. Software version 3.3 for Rapid Programmer devices, a personal digital assistant (PDA) based programmer supporting implantable pulse generators of the Eon, Genesis, and Renew family.
- Product code
- LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
- Reason for recall
- Product programmed with unapproved software was distributed.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- St. Jude Medical Neuromodulation contacted all affected customers by telephone between June 2, 2009 and June 5, 2009. Customers were given information about the product, problem and action to be taken. All affected devices were returned to the firm as instructed.
- Quantity in commerce
- 14 units
- Distribution
- Distributed in New Hampshire and Puerto Rico.
- Product codes and lots
- Serial numbers: 101565, 101544, 101555, 101548, 101561, 101551, 101563, 101546, 101556, 101553, 101554, 101558, 101545, and 101566