XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any...
Computerized Medical Systems Inc began a recall of XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any cancer patient for who external beam radiation... on . The reason FDA records is "XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2582-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any cancer patient for who external beam radiation therapy of brachytherapy has been prescribed.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values. If the algorithm is then changed to Monte Carlo, the calculation will proceed and the user is not warned the Material/Thickness values just entered are not used in the dose calculati
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Elekta CMS Software drafted a User Notice which explains the reason for correction, the clinical impact, and the workaround method. The User Notice was distributed to all affected sites in November 2010. The issue will be resolved in the XiO Release 4.70.00, expected to be available in July 2011.
- Quantity in commerce
- 556
- Distribution
- Worldwide Distribution - USA (nationwide, Washington, DC and Puerto Rico and the countries of Albania, Algeria, Australia, Austria, Bahamas, Belarus, Belgium, Brazil, Canada, Chile, China,Croatia, Cyrus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Phillipines,Poland, Portugal, Romoania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Suriname, Switzerland, Taiwan, Thailand, Turkey, Turmenistan, Ukraine, United Kingdom, and Vietnam
- Product codes and lots
- Release 4.50 and higher
FIRST ARTICLE / WEEKLY