King System , King LTSD Kit Size 4, non-sterile, REF KLTSD414, Manufacturer: King Systems Noblesville, IN...
King Systems Corp. began a recall of King System , King LTSD Kit Size 4, non-sterile, REF KLTSD414, Manufacturer: King Systems Noblesville, IN. Ventilation during anesthesia during procedures of... on . The reason FDA records is "The firms contract manufacturer reported that the products were below the specification for hardness. This could contribute to a higher incidence of problems during the insertion and final positioning of the airway in...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1878-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- King Systems Corporation.
- Product
- King System , King LTSD Kit Size 4, non-sterile, REF KLTSD414, Manufacturer: King Systems Noblesville, IN. Ventilation during anesthesia during procedures of short duration. An oropharyngeal airway is a device inserted through the mouth to provide a patent airway.
- Product code
- CAE Airway, Oropharyngeal, Anesthesiology
- Reason for recall
- The firms contract manufacturer reported that the products were below the specification for hardness. This could contribute to a higher incidence of problems during the insertion and final positioning of the airway in the patient. In rare instances the softness of the tube could cause the tube to fold over.
- Root cause tag
- supplier
- FDA root cause
- Other
- Action
- King Systems Corp. sent an URGENT DEVICE RECALL NOTICE letter dated January 24, 2011, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. The letters instructed users to cease all sales of the recalled product, to notify the consignees for any product that was further distributed, and to return all recalled product to King Systems. The notice did include a response form. For questions call 317-776-6823, ext 257.
- Quantity in commerce
- 138 cases 5 units per case, 690 units
- Distribution
- Worldwide Distribution - USA and Canada
- Product codes and lots
- IV715