Z-1228-2011

Thomas Medical Products, Acuity Bleed back Control Valve, Acuity Rotating Hemostasis Valve, Acuity Universal...

Merit Medical Systems, Inc. / initiated 2010-12-16 / Terminated / root cause: packaging

Thomas Medical Products Inc began a recall of Thomas Medical Products, Acuity Bleed back Control Valve, Acuity Rotating Hemostasis Valve, Acuity Universal Cutter, and Contamination Sleeve for use during... on . The reason FDA records is "Compromised sterility/weak or open seals". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1228-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Thomas Medical Products, Acuity Bleed back Control Valve, Acuity Rotating Hemostasis Valve, Acuity Universal Cutter, and Contamination Sleeve for use during vascular interventional procedures. Contamination Sleeve part no. CON-006-02; Acuity Bleedback Control Valve part no. CSP-001-00; Acuity Rotating Hemostasis Valve part no. CSP-002-01; Acuity Universal Cutter part no. CSP-003-00; Acuity Universal Cutter part no. CSP-003-01.
Product code
DQY Catheter, Percutaneous
Reason for recall
Compromised sterility/weak or open seals.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
GE Healthcare Thomas medical products Recall notification dated 16 December 2010 (GEHC Ref # 35003) was sent by letter via UPS Next Day Air and/or USPS certified mail.
Quantity in commerce
28,679 for all products
Distribution
Distribution Nationwide USA and Germany.
Product codes and lots
Contamination Sleeve part no. CON-006-02, lot no. S32943; Acuity Bleedback Control Valve part no. CSP-001-00, lot no. Z0000040; Acuity Rotating Hemostasis Valve part no. CSP-002-01, lot no. Z0000036; Acuity Universal Cutter part no. CSP-003-00, lot numbers Z0000038 and Z0000039; Acuity Universal Cutter part no. CSP-003-01, lot number Z0000054.