Z-1728-2011

Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162...

Qiagen / initiated 2011-01-12 / Terminated / root cause: software

QIAGEN Gaithersburg, Inc. began a recall of Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162, and as a component within a Qiagen PC, available... on . The reason FDA records is "The software component of a diagnostic medical device may be defective and could cause incorrect patient values". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1728-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Qiagen
Product
Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162, and as a component within a Qiagen PC, available as catalog # 5050-11140A. The software calculates and interprets assay results generated by the Digene Microplate Luminometer according to assay validation parameters. The software is used in conjunction with the following QIAGEN products: digene HC2 High-Risk HPV DNA Test [5199-1220 (one-plate); 5199-00016 (four-plate)] digene HC2 HPV DNA Test (5198-1220) digene HC2 CT/GC DNA Test (5130-1220) digene HC2 CT-ID DNA Test (5135-1220) digene HC2 GC-ID DNA Test (5140-1220) digene HC2 CT-GC Dual-ID DNA Test (5145-1220)
Product code
JJQ Colorimeter, Photometer, Spectrophotometer For Clinical Use
Reason for recall
The software component of a diagnostic medical device may be defective and could cause incorrect patient values.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Qiagen, sent a "Notice of Product Field Action" letter dated January 12, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue the process of running multiple assay protocols on a single microplate to eliminate the potential for inaccurate patient result reporting, and to complete and return the acknowledgment section via fax to QIAGEN 661-702-3854 or scan the document and email to [email protected]. The firm informed users that they will provide further detailed instructions for a workaround and software fix. A QIAGEN representative will contact the customers who indicated that they run multiple assays on a plate and are currently using the HC2 System Software Version 3.0 (Suite 4.0), to assist in the identification of those patients that may have been affected by the software defect. If you have any questions or concerns, please contact Qiagen Technical Services at 800-344-3631, Option 2.
Quantity in commerce
163
Distribution
The recalled product was sold to medical laboratories nationwide and in Canada.
Product codes and lots
hc2 User's Guide, Lots 5723098 and 5722169; and Qiagen embedded hc2 Software Suite 4.0, Lots Q50455 and Q50349.