Heart Sync Adult Radiolucent Electrode -Physio Control, Part T100LO-PHYSIO. Indicated for use in external...
Heart Sync, Inc began a recall of Heart Sync Adult Radiolucent Electrode -Physio Control, Part T100LO-PHYSIO. Indicated for use in external pacing, defibrillation and monitoring applications as... on . The reason FDA records is "The "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1951-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Graphic Controls Acquisition Corp.dba Nissha Medical Technologies
- Product
- Heart Sync Adult Radiolucent Electrode -Physio Control, Part T100LO-PHYSIO. Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only.
- Product code
- MLN Electrode, Electrocardiograph, Multi-Function
- Reason for recall
- The "leads out" connector had bent pins which would not allow connection to the Medtronic Physio Control Defibrillator cable during a product demonstration.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- The firm sent a recall notice to their distributors on 12/3/2008 asking them to quarantine product and return it to HeartSync. The letter stated that customers' current inventory would be exchanged for new product. Customers were to call 1-800-828-4681 to provide the necessary information.
- Quantity in commerce
- 50 (5 cases of 10 each)
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Number: Y111208-6.