Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic...

Integra LifeSciences Corporation / initiated 2011-01-12 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic treatment platform. Head Rings are used to... on . The reason FDA records is "Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1400-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic treatment platform. Head Rings are used to provide a reference frame for instrumentation used for precise spatial localization and treatment of physiologic targets for stereotactic neurosurgical procedures such as craniotomies, biopsies, functional neurosurgery, and radiation therapy. Head Rings are delivered to the user non-sterile, and are reusable.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach
Root cause tag
manufacturing process
FDA root cause
Manufacturing material removal
Action
Integra sent a Medical Device Field Corrective Action letter dated January 12, 2011, to all affected customers via Federal Express. Integra Sales Specialists were provided with the replacement part and written instructions on how to replace the affected intubation hoops from customer inventory identified in the traceability report. Upon completion of the field correction each sales specialist was instructed to return the affected hoop along with a completed copy the Recall Acknowledgement and Return Form. If additional information is required contact me at (609) 936-2485 or [email protected].
Quantity in commerce
8 units
Distribution
Worldwide Distribution - USA, Canada, and Australia
Product codes and lots
Lot/Serial Number(s): 0189463, 0189289, 0186642, 0184885