GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research
GE Healthcare, LLC began a recall of GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research on . The reason FDA records is "Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0833-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE eXplore Locus Micro CT cabinet x-ray system model RS9-80. CT Scanner for laboratory mice or other research.
- Product code
- RCE Cabinet X-Ray, Industrial
- Reason for recall
- Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011.
- Quantity in commerce
- 21
- Distribution
- Worldwide Distribution
- Product codes and lots
- Part Number 5134163.