Z-2569-2011

Monaco Radiation Treatment Planning Workstation. Monaco Release 2.04.00 and above. Used to create treatment...

Elekta Limited / initiated 2010-10-27 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of Monaco Radiation Treatment Planning Workstation. Monaco Release 2.04.00 and above. Used to create treatment plans for any cancer patient for who external beam... on . The reason FDA records is "Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2569-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco Radiation Treatment Planning Workstation. Monaco Release 2.04.00 and above. Used to create treatment plans for any cancer patient for who external beam intensity modulated radiation therapy (IMRT) has been prescribed.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.
Root cause tag
software
FDA root cause
Software design
Action
Computerized Medical Systems, Inc. sent a User Notice to the affected consignees on October 27, 2010, explaining the reason for correction, the clinical impact, and the workaround method. (VERIFY USE NOTICE DATE) A return postcard was included for the customer to confirm receipt of the User Notice. The firm intends for this issue to be corrected in Monaco Release 3.00 in March 2011.
Quantity in commerce
11
Distribution
Worldwide Distribution - USA including CA, IL, MN, NM, OH, NY, TX, and WA and the Netherlands.
Product codes and lots
Release 2.04 and higher