Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1521-2011

Vari Lase WireFiber Laser Fiber, RED 7137, Rx ONLY, Sterile EO, Vascular Solutions Inc., 6464 Sycamore Court...

Vascular Solutions LLC / initiated 2011-01-10 / Terminated / root cause: manufacturing process

Vascular Solutions, Inc. began a recall of Vari Lase WireFiber Laser Fiber, RED 7137, Rx ONLY, Sterile EO, Vascular Solutions Inc., 6464 Sycamore Court, Minneapolis, MN 55369 The VARI-LASE WireFiber is... on . The reason FDA records is "Investigations of a recent Device Experience Report has made us aware of a potential problem with our Vari-Lase WireFiber, Model 7137, manufactured with the following lot numbers: 5500898 and 550896. Recently it was...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1521-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vascular Solutions LLC
Product
Vari Lase WireFiber Laser Fiber, RED 7137, Rx ONLY, Sterile EO, Vascular Solutions Inc., 6464 Sycamore Court, Minneapolis, MN 55369 The VARI-LASE WireFiber is indicated for the treatment of varicose veins and varicosities associated with superficial reflux in the Great Saphenous Vein and for treatment of incompetence and reflux of superficial veins in the lower extremity.
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
Investigations of a recent Device Experience Report has made us aware of a potential problem with our Vari-Lase WireFiber, Model 7137, manufactured with the following lot numbers: 5500898 and 550896. Recently it was reported that the endovenous laser fiber cage and guidewire tip of the WireFiber became separated. It is possible that separation could occur during patient procedure and has the p
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Vascular Solutions sent an "Urgent Medical Device Recall" letter dated January 10, 2011. The letter described the issue and the product affected. Advised consignees to immediately check their inventory , place affected product in a secure location, and to complete and return the Customer Inventory Form. For questions regarding this recall call 763-656-4210.
Quantity in commerce
41
Distribution
Nationwide Distribution - USA including CT, MA, MN, NY, VA
Product codes and lots
Lot #'s 550896 and 550898