Z-1734-2011

Thermo Scientific HM 550 Cold Disinfection Cryostat, Microm International GmbH, Waldorf, Germany Tissue...

Thermo Fisher Scientific Oy / initiated 2010-09-23 / Terminated / root cause: design

Thermo Fisher Scientific began a recall of Thermo Scientific HM 550 Cold Disinfection Cryostat, Microm International GmbH, Waldorf, Germany Tissue processing equipment used to prepare human tissue... on . The reason FDA records is "Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfection efficacy". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1734-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Thermo Fisher Scientific Oy
Product
Thermo Scientific HM 550 Cold Disinfection Cryostat, Microm International GmbH, Waldorf, Germany Tissue processing equipment used to prepare human tissue specimens for diagnostic histological examination by processing specimens through the various stages of decalcifying, infiltrating, sectioning, and mounting on microscope slides.
Product code
IDP Microtome, Cryostat
Reason for recall
Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfection efficacy.
Root cause tag
design
FDA root cause
Device Design
Action
ThermoFisher sent Advisory Notice letters beginning on September 28, 2010, to all affected customers. The letters stated that the firm needed to conduct more comprehensive system validation data on the Cold-Disinfection sytem option available on the HM550-D cryostat to support the claim of complete bacterial, fungal and viral disinfection efficacy. The letter recommended that the Cold Disinfection option on the HM550-D not be used as the sole disinfection method on the HM550-D until the validation work is completed. The firm provided addtional safety protocols pertaining to the cryostat disinfection. Customers were instructed to complete and fax the Confirmation of Receipt Form to Technical Support, COLD-D Advisory Notice at (269) 372-2436.
Quantity in commerce
100 units
Distribution
Worldwide Distribution - USA including the states of LA, OH, FL, MD, PA, TX, MO, WI, KS, AL, CA, VA, MS, IL, NE, MA, and the countries of Australia, Belgium, China, Czech Republic, Denmark, Estonia, Finnland, France, Germany, Hong Kong, India, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russia, Spain, Sweden, Taiwan, and the UK.
Product codes and lots
Upgrade 37379, Upgrade 40963, Upgrade 41256, 47686, 47778, 47927, 47981, 48018, 48019, 48052, 48053, 48125, 48126, 48127, 48352, 48559, 46594, 47048, 47064, 47213, 47226, 47737, 47738, 47874, 47875, 48228, 48229, 48355, 48356, 46607, 46608, 48350, 46605, 47693, 48629, 48022, 48227, 48134, 46823, 46824, 48716, 43642.