Z-1862-2011

Xcelerate Specialty Patella Milling System Per File K2399; Catalog Number I-K2399PR00, I-K2399PR10; Stryker...

Howmedica Osteonics Corp / initiated 2011-01-27 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Xcelerate Specialty Patella Milling System Per File K2399; Catalog Number I-K2399PR00, I-K2399PR10; Stryker Howmedica Osteonics, Howmedica Osteonics Corp., 325... on . The reason FDA records is "Stryker Orthopaedics has become aware that the 10 mm on the bone remaining depth gage is incorrect thus allowing for the reamer to leave as little as 8.9mm (in a nominal condition) patella bone". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1862-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Xcelerate Specialty Patella Milling System Per File K2399; Catalog Number I-K2399PR00, I-K2399PR10; Stryker Howmedica Osteonics, Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The Xcelerate patella milling system (as detailed in surgical protocol LSPK35) was designed to prepare a patella to accept all Scorpio or Duracon patellar implants using either a resurfacing or recessed technique.
Product code
HTJ Gauge, Depth
Reason for recall
Stryker Orthopaedics has become aware that the 10 mm on the bone remaining depth gage is incorrect thus allowing for the reamer to leave as little as 8.9mm (in a nominal condition) patella bone.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker sent an Urgent Product Recall letter dated January 27, 2011, via FedEX for January 28, 2011, delivery with return receipt. The letter identified the product, the problem, and the action to be taken by the customer. Customers were insructed to fax back the attached Product Recall Acknowledgement Form within 5 days to 201-831-6069. Customers were also instructed to contact the hospitals in their territory that have the affected product to arrange return of the product. Product should be returned to the attention of: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey, 07430. The orange Product Remediation sticker should be attached to the return, indicating the Product Remediation # RA 2009-296. Mark the outer box with words "Product Recall." For questions regarding this recall call (201) 972-2100.
Quantity in commerce
52 units
Distribution
Nationwide Distribution
Product codes and lots
All lots