Z-1835-2011

Triathlon PKR Peg Drill, Catalog Number: 5650-4-512 (Size 1-2 Peg Drill); Catalog Number: 5650-4-536 (Size...

Howmedica Osteonics Corp / initiated 2011-02-09 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Triathlon PKR Peg Drill, Catalog Number: 5650-4-512 (Size 1-2 Peg Drill); Catalog Number: 5650-4-536 (Size 3-6 Peg Drill); Howmedica Osteonics Corp. 325... on . The reason FDA records is "Stryker Orthopaedics has identified a trend of complaints involving reports of the peg drill cold welding to the drill guide, causing the Peg Drill to cease functioning. If cold welding of these components occurs, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1835-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Triathlon PKR Peg Drill, Catalog Number: 5650-4-512 (Size 1-2 Peg Drill); Catalog Number: 5650-4-536 (Size 3-6 Peg Drill); Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 The Triathlon-PKR Knee System is indicated for primary partial knee resurfacing (PKR). This system utilizes two fixation pegs on the femoral component for fixation and alignment. To prepare for these features, a peg drill, drill guide, and drill through femoral trial are provided. The femoral Trial Drill Guide is attached to the Modular Handle and assembled into the holes on the articulation surface of the Femoral Trial. Using the appropriately sized Peg Drill (Small Drill for sizes 1-2, Large Drill for sizes 3-6), both holes are drilled. The Peg drill is advanced until the step on the drill makes contact with the front face of the Femoral Trial Drill Guide.
Reason for recall
Stryker Orthopaedics has identified a trend of complaints involving reports of the peg drill cold welding to the drill guide, causing the Peg Drill to cease functioning. If cold welding of these components occurs, the following potential patient and/or surgeon or surgical staff harms may results.
Root cause tag
manufacturing process
FDA root cause
Component change control
Action
Stryker sent Urgent Product Recall letters on February 9, 2011, and February 10, 2011, via Fed Ex. to all affected customers. The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete and fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For questions regarding this recall call 201-972-2100.
Quantity in commerce
5650-4-512- 754 units; 5650-4-536- 880 units
Distribution
Worldwide Distribution - USA , Sweden, Germany; France, Portugal, UK, Canada, Australia, and New Zealand,
Product codes and lots
Catalog Number 5650-4-512 Catalog Number 5650-4-536