Z-1615-2011

STA-Flex SUBTALAR SCREW, 10 mm x 16 mm, REF 982364, Sterile, Biomet Sports Medicine, 56 East Bell Drive...

Zimmer GmbH / initiated 2011-01-21 / Terminated

Biomet, Inc. began a recall of STA-Flex SUBTALAR SCREW, 10 mm x 16 mm, REF 982364, Sterile, Biomet Sports Medicine, 56 East Bell Drive, Warsaw, IN. Used to reduce excessive rearfoot... on . The reason FDA records is "The firm has initiated this recall following internal fatigue testing that shows possible initiation of fatigue cracking in implants under load". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1615-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
STA-Flex SUBTALAR SCREW, 10 mm x 16 mm, REF 982364, Sterile, Biomet Sports Medicine, 56 East Bell Drive, Warsaw, IN. Used to reduce excessive rearfoot pronation by blocking forward and downward displacement of the Talus, resulting in improved rearfoot alignment and foot function.
Product code
HWC Screw, Fixation, Bone
Reason for recall
The firm has initiated this recall following internal fatigue testing that shows possible initiation of fatigue cracking in implants under load.
FDA root cause
Pending
Action
The firm sent URGENT MEDICAL DEVICE RECALL NOTICES dated 1/21/2011 to distributors, implanting physicians, and medical facilities. The letter to the distributors and medical facilities states the recalled product should be immediately located and returned to Biomet. In addition, the "FAX Back Response Form" should be completed and returned to the number provided. If product has been further distributed, distributors are to contact hospital personnel via the enclosed "Dear Biomet Customer" notice. The letter to the implanting physicians informs the physician of the recall and lists the patients that were implanted with the device. The physician letter quotes from the surgical technique "the implant should also be removed if any postoperative complications occur". The letter to the implanting physicians does not recommend any action be taken with the devices that are implanted. The notice just makes the surgeons aware of the recall. Distributors and Physicians are to confirm receipt of the letter by contacting 800-348-9500, ext 3755 or 3756. Also, questions related to this matter should be directed to 574-371-3755 or 574-371-3756.
Quantity in commerce
254 all products
Distribution
Nationwide Distribution.
Product codes and lots
Lot Numbers: 384950, 879290, and 890870.