Z-3273-2011
Synthes Bone Marrow Aspiration System, Sterile, 11 gauge, 15cm needle (without side holes) Indicated for...
Synthes GmbH / initiated 2011-01-26 / Terminated / root cause: manufacturing process
Synthes USA (HQ), Inc. began a recall of Synthes Bone Marrow Aspiration System, Sterile, 11 gauge, 15cm needle (without side holes) Indicated for aspirating bone marrow on . The reason FDA records is "Firm became aware that some systems were released to stock without going through the sterilization process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3273-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Bone Marrow Aspiration System, Sterile, 11 gauge, 15cm needle (without side holes) Indicated for aspirating bone marrow
- Product code
- FMF Syringe, Piston
- Reason for recall
- Firm became aware that some systems were released to stock without going through the sterilization process.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Synthes (USA) sent an "URGENT MEDICAL DEVICE RECALL" letter dated February 4, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. Affected product removed from inventory will be replaced by Synthes. Contact the Synthes Spine Sale Consultant at 1-800-620-7025 ext.6883 for questions concerning this recall.
- Quantity in commerce
- 96 parts
- Distribution
- Nationwide Distribution-including the states of AR, CO, FL, GA, ID, LA, MA, MD, MO, MT, NY, NC, OH, SC, SD, TN, TX, and VA.
- Product codes and lots
- Part Number 710.150.99S, Lot number N999822