Z-1497-2011
High Viscosity Dermabond ProPen Topical Skin Adhesive 2-Octo Cyanoacrylate, Sterile, Rx only, Ethicon, Inc
Ethicon, Inc / initiated 2011-02-02 / Terminated / root cause: manufacturing process
Ethicon, Inc. began a recall of High Viscosity Dermabond ProPen Topical Skin Adhesive 2-Octo Cyanoacrylate, Sterile, Rx only, Ethicon, Inc on . The reason FDA records is "Some units of Dermabond Topical Skin Adhesive and Dermabond ProPen Topical Skin Adhesive contain discolored product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1497-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon, Inc
- Product
- High Viscosity Dermabond ProPen Topical Skin Adhesive 2-Octo Cyanoacrylate, Sterile, Rx only, Ethicon, Inc.
- Product code
- KGX Tape And Bandage, Adhesive
- Reason for recall
- Some units of Dermabond Topical Skin Adhesive and Dermabond ProPen Topical Skin Adhesive contain discolored product.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Ethicon sent their recall letters with business reply cards attached on February 2, 2011 via UPS overnight mail. Customers are informed of the issue and asked to return the attached business reply card.
- Quantity in commerce
- 725,722 each (total)
- Distribution
- Worldwide distribution, including USA, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Czech Repub, Chile, China, Denmark, Dominican Repub, Egypt, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, ireland, Israel, italy, Japan, Korea, Lithuania, Luxembourg, Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Palestine, Puerto Rico, Reunion, Russia, Senegal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Ukraine, and Venezuela.
- Product codes and lots
- Product code Product Code Lot # Product Expiration Date DPP6 CJP639 7/31/2012 DPP6 CJR201 7/31/2012 DPP6 CKB291 8/31/2012 DPP6 CKB882 8/31/2012 DPP6 CKE426 8/31/2012 DPP6 CKE893 8/31/2012 DPP6 CKP445 8/31/2012
FIRST ARTICLE / WEEKLY