Orthoralix 9200 DDE Panoramic and Cephalometric Dental X-Ray System; Extraoral x-ray unit with timer; Gendex...
Gendex Dental Systems began a recall of Orthoralix 9200 DDE Panoramic and Cephalometric Dental X-Ray System; Extraoral x-ray unit with timer; Gendex Dental Systems, Des Plaines, IL 60018; Part... on . The reason FDA records is "There is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2041-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gendex Corp.
- Product
- Orthoralix 9200 DDE Panoramic and Cephalometric Dental X-Ray System; Extraoral x-ray unit with timer; Gendex Dental Systems, Des Plaines, IL 60018; Part numbers 110-0199G2, 110-0199G4, 110-0234G1 and 110-0234G3. The intended use of the device is to be used as an extraoral source of x-rays for imaging of the dento-maxillo-facial area.
- Product code
- EHD Unit, X-Ray, Extraoral With Timer
- Reason for recall
- There is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obviou
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Gendex sent a recall letter to their accounts on February 4, 2011 via first class mail, stating that an anomaly in the Orthoralix firmware can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obvious these images cannot be used for diagnostic purposes. The letter stated that resending the image from the X-Ray system to the PC will correct the issue present in certain images sent during periods of high network traffic. Instructions were provided to download an Appendix to Orthoralix DDE User Manual that explains how to resend the images. If that does not resolve the issue, users were requested to contact Gendex Technical Support at 1-800-323-8029. Customers were asked to complete and return the enclosed Acknowledgement Form by fax to 215-997-5665 or e-mail: [email protected]. For questions regarding this recall call 800-323-8029.
- Quantity in commerce
- Total 6,835 units of all products
- Distribution
- Worldwide Distribution--United States, including Puerto Rico, Australia, Canada, Chile, Colombia, Germany, Hong Kong, India, Italy, Korea, Mexico, New Zealand, Peru, Singapore, Taiwan and Vietnam.
- Product codes and lots
- Part numbers 9869-401-02401, 9869-401-02501, 9869-401-02601, 9869-401-02701, 9869-401-02801, 9869-401-02901, and all serial numbers.