Z-1596-2011

Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number...

Abbott Medical / initiated 2010-10-27 / Terminated / root cause: design

Hospira Inc began a recall of Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One... on . The reason FDA records is "Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1596-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
Product code
FRN Pump, Infusion
Reason for recall
Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
Root cause tag
design
FDA root cause
Component design/selection
Action
Hospira Recall notification letters were sent by Federal Express on November 1, 2010.
Quantity in commerce
32,376 power cords
Distribution
Nationwide distribution and to Australia, Canada, and Malaysia.
Product codes and lots
Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser List number configurations: 16026-04-01/03/04, 51/52,. 53/54, 79/80, 81/82, 83/84/ 87/88; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88; 16026-13-25/26, 53/54; 16027-13-25/26, 53/54.