Z-1596-2011
Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number...
Abbott Medical / initiated 2010-10-27 / Terminated / root cause: design
Hospira Inc began a recall of Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One... on . The reason FDA records is "Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1596-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Hospira Power cord use on Symbiq One-Channel infuser and Symbiq Two-Channel infuser. Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser
- Product code
- FRN Pump, Infusion
- Reason for recall
- Complaints of broken, bent or missing prongs, charring, sparks, visible smoke, burnt smell have been reported on Symbiq AC Power cords.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Hospira Recall notification letters were sent by Federal Express on November 1, 2010.
- Quantity in commerce
- 32,376 power cords
- Distribution
- Nationwide distribution and to Australia, Canada, and Malaysia.
- Product codes and lots
- Power cord part number: HSP3306-E07. Used on List No. 16026 - Symbiq One Channel infuser; List No. 16027 - Symbiq Two-Channel infuser List number configurations: 16026-04-01/03/04, 51/52,. 53/54, 79/80, 81/82, 83/84/ 87/88; 16027-04-01/03/04, 51/52, 53/54, 79/80, 81/82, 83/84, 87/88; 16026-13-25/26, 53/54; 16027-13-25/26, 53/54.