Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1829-2011

Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker...

Stryker Corp. / initiated 2011-02-03 / Terminated / root cause: packaging

Stryker Instruments Division of Stryker Corporation began a recall of Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous... on . The reason FDA records is "The firm received five complaints that the stylet did not fit down the cannula. The investigation revealed that the incorrect probe assembly was packaged for two DeKompressor part numbers with the lot number of 10214012". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1829-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Interventional Pain 6" 13 G (2.4 mm) Percutaneous Discectomy Probe, REF 407-265, Sterile, Stryker Instruments, Kalamazoo, MI. The DeKompressor Percutaneous Discectomy Probe is intended for use in the apiration of disc material during percutaneous discectomies in the lumbar, thoracic, and cervical regions of the spine
Product code
HRX Arthroscope
Reason for recall
The firm received five complaints that the stylet did not fit down the cannula. The investigation revealed that the incorrect probe assembly was packaged for two DeKompressor part numbers with the lot number of 10214012.
Root cause tag
packaging
FDA root cause
Other
Action
Stryker Instruments Division of Stryker Corporation sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter dated February 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately check their inventory for any recalled product and quarantine any affected product. If the product was further distributed, customers were asked to forward the letter and the completed Reply Form to all affected locations. Customers were to be contacted to coordinate return of the affected product. Upon receipt of the product, a replacement would be issued to their account. For questions regarding this recall call 1-800-253-3210.
Quantity in commerce
44 units both products
Distribution
Worldwide Distribution - USA including the states of PA, NY, KS, NJ, TX, MI, OK, and FL, and the countries of Italy and Switzerland.
Product codes and lots
Lot: 10214012.