Z-1369-2011

Allura XPER FD 20 and and Allura CV20; Allura XPER FD 10 and Field Extension for XPER Cardiovascular System...

Philips North America LLC / initiated 2010-11-04 / Terminated

Philips Medical Systems North America Co. Phillips began a recall of Allura XPER FD 20 and and Allura CV20; Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura CV20. Angiographic X-ray systems with a... on . The reason FDA records is "Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1369-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Allura XPER FD 20 and and Allura CV20; Allura XPER FD 10 and Field Extension for XPER Cardiovascular System and Allura CV20. Angiographic X-ray systems with a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Adjustment of the micro-switches in the footswitch used on the Philips cardiovascular X-ray systems was not implemented. If it is programmed for exposure, it is possible that after pressing the pedal on the footswitch it does not release itself. The patient may be exposed to X-ray longer than intended.
FDA root cause
Pending
Action
On 11/5/10, Philips began sending the Electronic Product Radiation Warning notification letter, dated November 4, 2010, to their consignees. The consignees were informed that the middle pedal of the foot switch of the Allura FD10, FD20, CV20 systems, and the Allura Field Extension series does not not always become inactive (sticks) when pedal is released. The function of the middle pedal on the foot switch is freely programmable by the customer. If it is programmed for exposure and if the pedal does not release by itself, the patient will be exposed to X-ray longer than intended. The consignees were advised to check if their footswitch is possibly affected. Footswitch identification 452270000141 is noted. If the pedal does not release by itself press the red emergency stop button on the system. Consignees can call Philips Healthcare Call Center at 800-722-9377, #5 and Reference FCO 72200189.
Quantity in commerce
7 units in US
Distribution
Worldwide Distribution -- US, including states of CA, IL, MA, NC, and OC.
Product codes and lots
Site Numbers: 103331, 520481, 537743, 43721190, 43796227, 51126900, and 51333259.