Z-1833-2011

REF 32-485050 VANGUARD KNEE INSTRUMENTATION MICROPLASTY SLIDEX ANTERIOR/POSTERIOR, QTY: 1 The guide is used...

Zimmer GmbH / initiated 2011-02-10 / Terminated

Biomet, Inc. began a recall of REF 32-485050 VANGUARD KNEE INSTRUMENTATION MICROPLASTY SLIDEX ANTERIOR/POSTERIOR, QTY: 1 The guide is used for femoral sizing - Neutral or 3 degree femoral... on . The reason FDA records is "Biomet has initiated this action due to the scale on the stylus of the Vanguard MP Slldex AP Sizer being incorrect for this assembly resulting in the scale being out of position". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1833-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
REF 32-485050 VANGUARD KNEE INSTRUMENTATION MICROPLASTY SLIDEX ANTERIOR/POSTERIOR, QTY: 1 The guide is used for femoral sizing - Neutral or 3 degree femoral rotation feet are attached to the slidex AP sizing guide.
Product code
FZX Guide, Surgical, Instrument
Reason for recall
Biomet has initiated this action due to the scale on the stylus of the Vanguard MP Slldex AP Sizer being incorrect for this assembly resulting in the scale being out of position.
FDA root cause
Process change control
Action
The firm, Biomet, sent a "FIELD SAFETY NOTICE" dated February 11, 2011 to all customers (heads of Orthopaedic Departments/Operating Departments /Sterile Services Departments/Procurement /Supplies/ Risk Management). The notice described the product, problem and actions to be taken. The customers were instructed to immediately locate and discontinue use of the product; return to Biomet or their local Biomet distributor; ensure that the operating staff are made aware of this issue, and complete and return the FAX BACK RESPONSE FORM to Biomet UK Ltd or their local Biomet distributor via fax at +44 (0) 1656 645454. If you have any questions regarding this communication, please contact the Regulatory & Compliance Manager at +44 (0) 1793 645317 or email: [email protected].
Quantity in commerce
17
Distribution
Foreign only; UK, Malta, The Netherlands, Italy, Turkey, Switzerland.
Product codes and lots
LOT 32-485050 LOT NO. SA10 81254