Z-1882-2011

iSite PACS with iSite ViewForum Applications, R7.4V6 L1 L2, L3, Versions: 3.6.96, 3.6.101, 3.6.110, 4.1.96...

Philips North America LLC / initiated 2011-02-23 / Terminated / root cause: software

Philips Medical Systems began a recall of iSite PACS with iSite ViewForum Applications, R7.4V6 L1 L2, L3, Versions: 3.6.96, 3.6.101, 3.6.110, 4.1.96, 4.1.101, and 4.1.110 ; Product is manufactured and... on . The reason FDA records is "A software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1882-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
iSite PACS with iSite ViewForum Applications, R7.4V6 L1 L2, L3, Versions: 3.6.96, 3.6.101, 3.6.110, 4.1.96, 4.1.101, and 4.1.110 ; Product is manufactured and distributed by Philips Medical Systems, Foster City, CA ViewForum Applications: iSite ViewForum Applications is an image processing workstation software package designed to run on standard PC hardware. The hardware required is made up of "off-the-shelf" standard computer components. iSite ViewForum Applications software receives image data from medical scanning devices, such as CT and MRI, or from image archives. iSite ViewForum Applications performs viewing, image manipulation, communication, printing, and quantification of images.
Product code
LLC
Reason for recall
A software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements which could cause an incorrect diagnosis and incorrect treatment plan.
Root cause tag
software
FDA root cause
Software design
Action
Philips sent an Urgent Field Safety Notice letter dated February 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to assure that all colleagues were aware of the problem. When using the Volume Vision software, check the correctness of measurements (graphics and numerical values) prior to saving a captured image containing measurements. Save captured images containing measurements immediately after creating them and before altering the measurements in the original image view. For any subsequent measurements, create a new capture and save it immediately. For further information or support concerning this issues, customers should contact their local Philips representative at +1-877-328-2808.
Distribution
Worldwide Distribution.
Product codes and lots
All units of these iSite PACS.