Z-1678-2011

2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion

Animas Corp. / initiated 2011-02-25 / Terminated / root cause: manufacturing process

Animas Corporation began a recall of 2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion on . The reason FDA records is "Action is being undertaken to address the occurrence of adverse events associated with leaking cartridges. Firm became aware of this issue through customer complaints from pump users reporting the cartridges were leaking". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1678-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Animas Corp.
Product
2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion
Product code
LZG Pump, Infusion, Insulin
Reason for recall
Action is being undertaken to address the occurrence of adverse events associated with leaking cartridges. Firm became aware of this issue through customer complaints from pump users reporting the cartridges were leaking.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Animas Corporation sent an Urgent Recall letter dated February 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Change their cartridge immediately using the instructions provided. Return all of the unused affected product using the enclosed box. Shipments must be returned using the United States Postal Service. For technical questions call Animas Customer Support at 1-855-254-5668 or call the toll free number at 1-8777-937-7867. For any questions about returning the product call 1-877-280-2339. Customers were instructed to notify the firm by logging on to the following website: www.animasnotification.com or call the Call Center at 1-877-280-2339 if they have no product to return.
Quantity in commerce
384,180
Distribution
Worldwide Distribution - USA (nationwide) and France.
Product codes and lots
Lot numbers -- US: B201575, B201576, B201581, B201582, Exp. 7/31/2012; B201583, Exp. 8/31/2012; FRANCE: B201580, Exp. 7/31/2012.