2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion
Animas Corporation began a recall of 2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion on . The reason FDA records is "Action is being undertaken to address the occurrence of adverse events associated with leaking cartridges. Firm became aware of this issue through customer complaints from pump users reporting the cartridges were leaking". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1678-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Animas Corp.
- Product
- 2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion
- Product code
- LZG Pump, Infusion, Insulin
- Reason for recall
- Action is being undertaken to address the occurrence of adverse events associated with leaking cartridges. Firm became aware of this issue through customer complaints from pump users reporting the cartridges were leaking.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Animas Corporation sent an Urgent Recall letter dated February 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Change their cartridge immediately using the instructions provided. Return all of the unused affected product using the enclosed box. Shipments must be returned using the United States Postal Service. For technical questions call Animas Customer Support at 1-855-254-5668 or call the toll free number at 1-8777-937-7867. For any questions about returning the product call 1-877-280-2339. Customers were instructed to notify the firm by logging on to the following website: www.animasnotification.com or call the Call Center at 1-877-280-2339 if they have no product to return.
- Quantity in commerce
- 384,180
- Distribution
- Worldwide Distribution - USA (nationwide) and France.
- Product codes and lots
- Lot numbers -- US: B201575, B201576, B201581, B201582, Exp. 7/31/2012; B201583, Exp. 8/31/2012; FRANCE: B201580, Exp. 7/31/2012.