Case Label: REORDER NO. 6102 STERILE WELCON BRAND Standard Irrigation Tray Thumb Control Ring Piston Syringe...
Nurse Assist, Inc began a recall of Case Label: REORDER NO. 6102 STERILE WELCON BRAND Standard Irrigation Tray Thumb Control Ring Piston Syringe 20 ea/Box nurse assist incorporated 800-649-6800... on . The reason FDA records is "The irrigation procedure trays contain Triad Group Alcohol Prep Pads recalled by Triad Group because they have the potential to be contaminated with an objectionable organism". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2252-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nurse Assist, LLC
- Product
- Case Label: REORDER NO. 6102 STERILE WELCON BRAND Standard Irrigation Tray Thumb Control Ring Piston Syringe 20 ea/Box nurse assist incorporated 800-649-6800 3400 Northern Cross Blvd. Fort Worth, TX 76137 www.nurseassist.com Product Label: REORDER NO. 6102 STERILE UNLESS OPENED OR DAMAGED nurse assist incorporated A6000, R1 WELCON BRAND Standard Irrigation Tray With Piston Syringe SINGLE USE 800-649-6800 3400 Northern Cross Blvd. Fort Worth, TX 76137 www.nurseassist.com CONTENTS: Underpad 500 mL Graduate 1200 mL Graduated Tray 60cc Thumb Ring Piston Syringe Tip Protector Alcohol Wipe Intended use: The trays are irrigation trays for catheters
- Product code
- FCM Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)
- Reason for recall
- The irrigation procedure trays contain Triad Group Alcohol Prep Pads recalled by Triad Group because they have the potential to be contaminated with an objectionable organism.
- FDA root cause
- Other
- Action
- The firm, Nurse Assist Incorporated, sent a letter dated January 20, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory immediately; quarantine the product; identify their customers and notify them at once about the product recall (Nurse Assist will provide bright labels to affix to your affected inventory); provide their customers with labels as well, and complete and return the attached RECALL ACKNOWLEGEMENT form via fax to: 817-231-1500 or mail to: Nurse Assist, Inc., 3400 Northern Cross Blvd., Forth Worth, TX 76137. Only the product and lots identified on the attached list are affected. If you have any questions or to request labels, please call Nurse Assist Customer Service Monday through Friday, from 8:00 am to 5:00 pm Central Time at 1-800-649-6800.
- Quantity in commerce
- 7,288 Cases
- Distribution
- Worldwide distribution: USA (nationwide) and countries including: Canada, Dubai and Saudi Arabia.
- Product codes and lots
- Lot Codes: 808244, 811011, 812145, 902102, 903108, 904077, 905023, 906198, 907079, 907279, 908130, 909284, 910265, 912282, 1001224, 1002216, 1003119, 1004169, 1005224, and 1006295.