Z-1910-2011
SPIDER ADVANCED Surgical Platform The SPIDER is a single port, single incision device to facilitate...
Asensus Surgical, Inc. / initiated 2011-02-25 / Terminated
Transenterix, Inc began a recall of SPIDER ADVANCED Surgical Platform The SPIDER is a single port, single incision device to facilitate multi-instrument access during laparoscopic surgical... on . The reason FDA records is "Plastic shards may break off and fall into the abdominal space which may require retrieval during surgery". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1910-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Asensus Surgical, Inc.
- Product
- SPIDER ADVANCED Surgical Platform The SPIDER is a single port, single incision device to facilitate multi-instrument access during laparoscopic surgical procedures. The SPIDER device is provided pre-sterilized and is a single use device.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Plastic shards may break off and fall into the abdominal space which may require retrieval during surgery.
- FDA root cause
- Other
- Action
- TransEnterix sent an Urgent Device Recall letter dated February 25, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the lot numbers affected, complete the attached inventory reconciliation form and return the form, along with any affected product to TransEnterix using the enclosed shipping label. Customers were asked to return the inventory form even if they had none of the lot numbers affected in their inventory. Their sales representative and customer service will work with them to replace the product in full at not cost to them. For questions regarding this recall contact TransEnterix Customer Care at 1-888-879-4111.
- Quantity in commerce
- 92 units
- Distribution
- Nationwide Distribution including AZ, FL, NC, NY, OH, OR, SC, TX, VA and WA
- Product codes and lots
- Model number: 9000021, Lot numbers: TRX056201001, TRX061201002, TRX071201001, TRX086201003.