Z-1886-2011

Harmony LA/LED Monitor Arm Adapter. Model Number: LB61 The Harmony LA/LED Monitor Arm Adapter is an accessory...

STERIS Corporation / initiated 2011-02-18 / Terminated / root cause: labeling and UDI

Steris Corporation began a recall of Harmony LA/LED Monitor Arm Adapter. Model Number: LB61 The Harmony LA/LED Monitor Arm Adapter is an accessory to the Harmony LA and LED Surgical Lighting... on . The reason FDA records is "The arm adapter does not contain labelling specifying the applicable weight and moment limitations of the arm adapter". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1886-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
Harmony LA/LED Monitor Arm Adapter. Model Number: LB61 The Harmony LA/LED Monitor Arm Adapter is an accessory to the Harmony LA and LED Surgical Lighting System and is attached to the end of the monitor suspension arm to allow Customers to attach other medical devices, such as non-STERIS monitors, camera or other compatible items in place of the STERIS monitor. The Harmony Surgical Lighting and Media System is a family of variable pattern, variable intensity surgical lights designed to provide visible immunization of the surgical field or the patient and to provide audio-visual procedural support for the O.R. staff. The device is intended to be used by surgeons and other medical care practitioners in a surgical setting
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
The arm adapter does not contain labelling specifying the applicable weight and moment limitations of the arm adapter
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Steris Corporation sent an Urgent Field Correction Notice letter dated February 18, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. A Steris representative would contact customers to arrange for replacement of the product. For questions regarding the firms visit to their facility, customers were instructed to contact Steris Field Service Dispatch at 1-800-333-8828. For any questions regarding this matter call 440-392-7705.
Quantity in commerce
14 Units
Distribution
Worldwide Distribution - USA including CO, IL, IN, MO, NC, NV, NY and OH and the countrries of Canada, China, and Australia.
Product codes and lots
Model Number: LB61