Z-1894-2011

BD Beaver XStar Safety 2.5mm Crescent Knife; REF 378234 Crescent blades are used in a variety of ophthalmic...

Becton Dickinson & Co / initiated 2011-02-28 / Terminated / root cause: manufacturing process

Beaver-Visitec International Inc. began a recall of BD Beaver XStar Safety 2.5mm Crescent Knife; REF 378234 Crescent blades are used in a variety of ophthalmic surgery. Product is a cataract surgery knife... on . The reason FDA records is "Loose blades in the handle of knives due to uncured epoxy". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1894-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Beaver XStar Safety 2.5mm Crescent Knife; REF 378234 Crescent blades are used in a variety of ophthalmic surgery. Product is a cataract surgery knife designed for optimal tunnel architecture.
Product code
H--NN
Reason for recall
Loose blades in the handle of knives due to uncured epoxy.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Beaver Visitec International, Inc.sent an Urgent Recall Notification letter dated February 25, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return all unused product and to complete a Customer Response Form (to be returned via fax to 866-906-4304. For questions regarding this recall call 781-906-7917.
Quantity in commerce
1680 units (1130 domestic, 550 foreign)
Distribution
Worldwide Distribution - USA including AZ, MA, MI, MO, NC, NJ, NY, OH, OR, PA, RI, TN, TX, VA and the country of China
Product codes and lots
Lot/serial number: 0333798