Z-1959-2011

Axiom Aristos FX and FX Plus System

Siemens Medical Solutions USA, Inc. / initiated 2011-02-17 / Terminated

Siemens Medical Solutions USA, Inc began a recall of Axiom Aristos FX and FX Plus System Intended use: Stationary x-ray system on . The reason FDA records is "During product monitoring, firm discovered a possible hazard to patients when using the detector alpha rotation of all Axiom Aristos FX and FX Plus systems with serial numbers up to and including 1216. When using the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1959-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Axiom Aristos FX and FX Plus System Intended use: Stationary x-ray system
Product code
KPR System, X-Ray, Stationary
Reason for recall
During product monitoring, firm discovered a possible hazard to patients when using the detector alpha rotation of all Axiom Aristos FX and FX Plus systems with serial numbers up to and including 1216. When using the detector alpha rotation, the detector may descend unexpectedly during movement or in vertical or table mode. If this occurs in vertical modes, the possibility exists that the detect
FDA root cause
Other
Action
The firm, SIEMENS, sent a "Safety Advisory Notice" dated February 15, 2011 to the affected customers via Update Instructions XP016/10/S on February 17, 2011. The notice described the product, problem and actions to be taken. The customers were instructed to immediately instruct their personnel accordingly; ensure that the safety advisory is placed in the systems's instructions for use; if they have sold this device/equipment and it is no longer in their possession, forward this safety notice to the new owner of the device/equipment; inform SIEMENS about the new owner of the device/equipment, and complete and return the Acknowledgement of receipt form. SIEMENS is preparing a modification of AXIOM Aristos FX and FX Plus that will resolve this potential malfunction. The field modification will be available from February 2011 and is expected to be carried out May 31, 2011. If you have any questions, call 09131-18-9908 or +1-888-826-9702.
Quantity in commerce
101
Distribution
Nationwide Distribution: USA including states of: AZ, AK, CA, CO, FL, GA. IL, IN, IA, KS, KY, ME, MI, MN, MS, MO, NE, NV, NJ, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, and WI.
Product codes and lots
Model numbers 10093864 and 7414803 -- Serial numbers up to and including 1216.