Z-2137-2011

NX is the radiographer's image identification and quality control tool. It offers a broad array of features...

AGFA Healthcare Corp. / initiated 2011-03-04 / Terminated

AGFA Corp. began a recall of NX is the radiographer's image identification and quality control tool. It offers a broad array of features developed especially to meet the needs of the... on . The reason FDA records is "When a radiographer is working on two open studies on an NX workstation and, in specific situations, a problem of image/annotation mix-up can occur on the NX system". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2137-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
NX is the radiographer's image identification and quality control tool. It offers a broad array of features developed especially to meet the needs of the radiographer. Standard tasks are completed quickly and effortlessly with a simple touch screen. In-room integration provides a more convenient workflow for the radiographer throughout the imaging process. NX communicates seamlessly with the hospital's picture archiving and communication system (PACS), radiology information system (RIS) and hospital information system (HIS). It is DICOM compliant and meets IHE guidelines.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
When a radiographer is working on two open studies on an NX workstation and, in specific situations, a problem of image/annotation mix-up can occur on the NX system.
FDA root cause
Other
Action
AGFA Corp. sent an "URGENT FIEL SAFETY NOTICE" letter via FED-EX to the customers on March 4, 2011. The letter describes the safety alert and mitigation. Customers were instructed to complete and return the Urgent Field Safety Notice Feedback Form by fax to 864-421-1664. For questions regarding this recall call 864-421-1815.
Quantity in commerce
775 units
Distribution
Worldwide Distribution - USA including CA, GA, MS, ND, NM, NY, OK, PA, SC, SD, TX, and VA and the country of Canada
Product codes and lots
Software Versions: 2.0.8100, 3.0.8100, 2.0.8200, 3.0.8200, 2.0.8300, 3.0.8300, 2.0.8400, 3.0.8400, 2.0.8500, and 3.0.8500