Z-2175-2011

Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information...

Philips North America LLC / initiated 2011-03-16 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information Portfolio Release D.03; 865209 IntelliVue... on . The reason FDA records is "Edits to an order can result in unintended scheduled interventions and a patient may receive additional interventions or medication doses that are not intended". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2175-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips IntelliVue Clinical Information Portfolio software as follows: 865047 IntelliVue Clinical Information Portfolio Release D.03; 865209 IntelliVue Clinical Information Portfolio Release E.0 and E.01; 865300 IntelliVue Clinical Information Portfolio NICU Link
Product code
NSX
Reason for recall
Edits to an order can result in unintended scheduled interventions and a patient may receive additional interventions or medication doses that are not intended.
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare notified accounts on 3/16/11 via the Urgent Medical Device Correction Notification. Customer notifications in the United States will be delivered via UPS. Customer notifications outside the US will be managed by Philips representatives in each affected geography. Customers will be asked to follow the guidelines in the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. All customers potentially affected by this issue will receive the software correction as part of the field action.
Quantity in commerce
387 units
Distribution
Worldwide distribution: USA, Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Kenya, Latvia, Malaysia, Netherlands, Norway, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
Product codes and lots
versions D.03, E.0, or E.01 of the software