Z-2015-2011
CELL-DYN Shear Valve Assembly used on CELL-DYN 3700SL Analyzer and CELL-DYN 3700SL Analyzer (refurb). Abbott...
Abbott Medical / initiated 2011-03-15 / Terminated
Abbott Laboratories began a recall of CELL-DYN Shear Valve Assembly used on CELL-DYN 3700SL Analyzer and CELL-DYN 3700SL Analyzer (refurb). Abbott Diagnostics Division, Santa Clara, CA 05054... on . The reason FDA records is "The Nylon Washer has an outer diameter which is undersized resulting in the inability to generate patient results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2015-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- CELL-DYN Shear Valve Assembly used on CELL-DYN 3700SL Analyzer and CELL-DYN 3700SL Analyzer (refurb). Abbott Diagnostics Division, Santa Clara, CA 05054 Designed for in vitro diagnostic use in clinical laboratories.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The Nylon Washer has an outer diameter which is undersized resulting in the inability to generate patient results.
- FDA root cause
- Pending
- Action
- Product Correction letters were sent on March 15, 2011. via Federal Express. The letter identified the affected product, explained the problem, the patient impact and the necessary action that were to be taken. Customers were to review the letter and inform their Laboratory Manager of the pending action.; complete and return the Customer Reply Form; and save a copy of the letter for their records. If there were any questions, customers were to contact Customer Support at 1-877-4ABBOTT or their local hematology customer support representative for customers outside of the US.
- Quantity in commerce
- 174 total, all varieties
- Distribution
- Worldwide Distribution -- AZ, CA, FL, IA, ID, IN, MA, MI, MN, MS, ND, NY, OK, OR, SC, SD, TX, WI, WY and Germany, Singapore, Mexico, China, South Korea, Argentina, Japan, Brazil, Canada, Australia, and Chile.
- Product codes and lots
- List Number 02H31-01; lot numbers: 23444AK, 23445AK, 23446AK, 23448AK to 23456AK, 23461AK, 20576AK, 21408AK, 21426AK, 21515AK, 21651AK, 21694AK, 271726AK, 21955AK, 22145AK, 22229AK, 22458AK, 22732AK, 22760AK, 23198AK, 23209AK, 23210AK, 23236AK, 22881AK, 19966AK, 20880AK, 20924AK, 21051AK, 21522AK, and 22650AK. Refurbished list number: 02H31-03; lot numbers: 20074AK96, 20334AK96, 20416AK96, 20751AK96, 20985AK96, 21167AK96, 21368AK96, 21610AK96, 21753AK96, 21842AK96, 21885AK96, 21934AK96, 22142AK96. 22340AK96, 22346AK96, 22380AK96, 22501AK96, 23056AK96, 23145AK96, 23457AK96, 23458AK96, 23459AK96, 23460AK96, and 23462AK96.