Z-2408-2011

XiO Radiation Treatment Planning System, XiO Release 1.0.2 and above. Used to create treatment plans for any...

Elekta Limited / initiated 2010-09-10 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of XiO Radiation Treatment Planning System, XiO Release 1.0.2 and above. Used to create treatment plans for any cancer patient for who external beam radiation... on . The reason FDA records is "When RT Plans are exported from XiO and RT Structure Sets and Images are exported from Monaco, the DICOM iso-center coordinates may not be accurate". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2408-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
XiO Radiation Treatment Planning System, XiO Release 1.0.2 and above. Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.
Product code
muj
Reason for recall
When RT Plans are exported from XiO and RT Structure Sets and Images are exported from Monaco, the DICOM iso-center coordinates may not be accurate.
Root cause tag
software
FDA root cause
Software design
Action
Elekta / Computerized Medical Systems Inc began mailing an Important Safety Notice to all affected customers in September 2010. The notice identified the product, the problem, and the action to be taken by the customer. Customers were advised not to mix treatment plan information from multiple planning systems. Customers were asked to return the postcard as proof of receipt and their understanding of the Safety Notice. For questions regarding this recall call 408-830-8000.
Quantity in commerce
1377
Distribution
Worldwide Distribution - USA (nationwide and Puerto Rico) and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, .Netherlands, New Zealand, Nicaragua, North Korea, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Suriname, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Turkey, Turkmenistan, Ukraine, United Kingdom, Uruguay, Venezuela, Yemen, and Zimbabwe.
Product codes and lots
XiO Release 1.0.2 and above