Z-2577-2011

XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for...

Elekta Limited / initiated 2008-04-08 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for any cancer patient for who external beam... on . The reason FDA records is "XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2577-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
XiO Radiation Treatment Planning System, XiO Release 4.0.0 through 4.40.00 Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
XiO: When the patient is positioned with their feet toward the CT gantry, the coordinates exported to the LAP IsoMark laser positioning system are not correct.
Root cause tag
software
FDA root cause
Software design
Action
The firm, CMS, Inc., sent a "XiO Customer Advisory" letter dated April 8, 2008 to all affected customers. The letter describes the product, the problem, and the actions to be taken ( workaround). The letter also states the problem will be resolved in a future release. No release number or date was given. If you have any questions, please call 408-380-8023.
Quantity in commerce
75
Distribution
Worldwide distribution: USA (nationwide) and countries of: China, France, Germany, Japan, Mongolia, Netherlands, Nicaragua, Poland, Russia, South Korea, Switzerland, and Turkey.
Product codes and lots
XiO Release 4.0.0 through 4.40.00