Monaco Radiation Treatment Planning System, Monaco Release 1.0.0 through 1.0.2. Used to create treatment...

Elekta Limited / initiated 2008-06-30 / Terminated / root cause: software

Computerized Medical Systems Inc began a recall of Monaco Radiation Treatment Planning System, Monaco Release 1.0.0 through 1.0.2. Used to create treatment plans for any cancer patient for who external beam... on . The reason FDA records is "Monaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2560-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco Radiation Treatment Planning System, Monaco Release 1.0.0 through 1.0.2. Used to create treatment plans for any cancer patient for who external beam intensity modulated radiation therapy (IMRT) has been prescribed.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Monaco: If the user selects Monte Carlo Photon as the Secondary Algorithm (to be used for final dose calculation through segments), and selects SKIP Forward button at any time during stage 2 optimization, the system may only use the pencil beam algorithm for final computation of dose even though Monte Carlo was selected. Other than specific output to the console window, there will be no other in
Root cause tag
software
FDA root cause
Software design
Action
Computerized Medical Systems, Inc. sent a "MONACO CUSTOMER ADVISORY" letter dated June 2008 to all affected customers. The letter describes the product, problem, and the action to be taken. The customer was instructed to use caution when using the Skip Forward button in stage 2 according to written instructions provided in the letter. This defect exists in Monaco Release 1.0.0 through 1.0.2 and will be resolved as soon as possible. Note: The issue was resolved in Monaco Release 2.0.1 and above. Questions should be directed to the manufacturer's representative telephone number at 408-830-8023.
Quantity in commerce
47
Distribution
Worldwide Distribution-- USA (nationwide) including the states of IL, MI, MO, OR, SC, TX, VA, and Puerto Rico and the countries of Canada, China, Cyprus, Estonia, France, Germany, Greece, Italy, Japan, Netherlands, Philippines, Poland, Portugal, Slovenia, South Korea, Spain, and United Kingdom.
Product codes and lots
XiO Release 1.0.0 through 1.0.2 All lot or serial numbers