The 4D Integrated Treatment Console; Model Number: H51; Reference: 4DITC versions 8.6 and below; File...
Varian Medical Systems, Inc. Oncology Systems began a recall of The 4D Integrated Treatment Console; Model Number: H51; Reference: 4DITC versions 8.6 and below; File Reference: CP-2007-0288; Product is manufactured and... on . The reason FDA records is "A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2119-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- The 4D Integrated Treatment Console; Model Number: H51; Reference: 4DITC versions 8.6 and below; File Reference: CP-2007-0288; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The 4D Integrated Treatment Console function is designed to assist the operator of radiation therapy device in providing accurate treatment setups for each patient by monitoring setup parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Varian Medical Systems sent an Urgent Medical Device Correction Letter on February 28, 2011, to all affected customers. Varian has created the procedure described in CTB-GE-579 to inform users on proper interfacing of the demountable BrainLAB m3 on the Varian Clinac. The procedure ( CTB-GE-579) is enclosed and posted to the customer site: http://www.MyVarian.com. This design fix in 4DITC v10.1 or later will be provided to all affected sites. If you require further clarification, please feel free to contact your local Varian Customer Support District or Regional Manager. USA and Canada - 1-888-VARIANS (888-827-4265) Europe - + 41 41 749 8844
- Quantity in commerce
- 17
- Distribution
- Class 2 Recall - Worldwide Distribution -- USA including states of ; CA, NV, NJ, GA, IL, and TX and countries of Australasia, Belgium, Brazil, China P.R., Denmark, France, Germany, Germany ( BrachyTherapy) Hong Kong, India Pvt. Ltd, Italy, Japan, Middle East, Netherlands, Russia, Spain, Switzerland and United Kingdom.
- Product codes and lots
- Affected Serial Numbers: H510378, H510392, H510712, H510844, H511781, H512506, H512547, H512645, H513143, H513196, H513460, H514034, H514385, H514468, H514759, H514941, H511891