Z-2051-2011

ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound...

Siemens Medical Solutions USA, Inc. / initiated 2011-03-15 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following... on . The reason FDA records is "The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2051-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following applications: Fetal, abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal and Peripheral Vascular Applications.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued a Customer Safety Advisory Notification letter to all customers on March 15, 2011, regarding a software error on the ACUSON S2000 ultrasound systems affected measurement . They are in the process of releasing a software update that will correct this issue. Siemens recommends reviewing the patient report at the end of each study to confirm that the report contains only measurements relevant to the exam just performed. No patient injury has been reported. Further questions regarding this recall please call (650) 694-5398.
Quantity in commerce
1533 units
Distribution
Worldwide Distribution including USA, states of AZ, VA, CA, NE, IL, MN, TN, FL, NM, and DC
Product codes and lots
Model Number 10041461-ACUSON S2000. Serial numbers to follow