Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1903-2011

REF 906764, CURVTEK EYE NEEDLE, 12 MM LARGE, STAINLESS STEEL GAMMA STERILIZED, PKG/3, BIOMET SPORTS MEDICINE...

Zimmer GmbH / initiated 2011-03-10 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of REF 906764, CURVTEK EYE NEEDLE, 12 MM LARGE, STAINLESS STEEL GAMMA STERILIZED, PKG/3, BIOMET SPORTS MEDICINE, 56 EAST BELL DRIVE, P.O. BOX 587, WARSAW, IN... on . The reason FDA records is "The firm initiated a recall due to the discovery that the devices were not sterilized, as required, prior to distribution". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1903-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
REF 906764, CURVTEK EYE NEEDLE, 12 MM LARGE, STAINLESS STEEL GAMMA STERILIZED, PKG/3, BIOMET SPORTS MEDICINE, 56 EAST BELL DRIVE, P.O. BOX 587, WARSAW, IN 46581 USA, STERILE After the Curvtek device is used to drill a "C" shaped tunnel through the patient's bone, the device is used to pass suture through the tunnel so that soft tissue may be tied down directly onto the bone.
Product code
GAB Needle, Suturing, Disposable
Reason for recall
The firm initiated a recall due to the discovery that the devices were not sterilized, as required, prior to distribution.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, BIOMET, sent an "URGENT MEDICAL DEVICE RECALL NOTICE" letter dated March 10, 2011 to all customers. The letter described the product, problem and actions to be taken. The letter noted that -Use of a non-sterile Curvtek Eye Needle may lead to patient infection, possibly requiring a revision procedure. The customers were instructed to immediately locate, discontinue use of the product , remove the product from circulation and return the product to Biomet; carefully follow the instructions on the enclosed FAX BACK RESPONSE FORM; complete and return the response form via fax to 574-372-1683 prior to return of product; use priority carrier for shipment, and please confirm receipt of this notice by calling 800-348-9500, extension 3755 or 3756. The firm also informed physicians that had previously utilized the device to monitor patients for infection and provide treatment as appropriate. Questions related to this notice should be directed to 574-371-3755 or 574-371-3756, Monday through Friday, 8am to 5pm.
Quantity in commerce
35
Distribution
Worldwide distribution: USA including states of: IN and KY; and countries of: Canada and The Netherlands.
Product codes and lots
Part Number 906764, Lots 748580, 748730 and 932650.