Case Label: REORDER NO. 7402 (STERILE) WELCON BRAND URETHRAL CATHETER KIT (Female) nurse assist incorporated...
Nurse Assist, Inc began a recall of Case Label: REORDER NO. 7402 (STERILE) WELCON BRAND URETHRAL CATHETER KIT (Female) nurse assist incorporated 25 ea/Box 800-649-6800 3400 Northern Cross Blvd... on . The reason FDA records is "The trays contain Triad Group Sterile Lubricating Jelly packs recalled by Triad Group because of the possibility of inadequate sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-2277-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nurse Assist, LLC
- Product
- Case Label: REORDER NO. 7402 (STERILE) WELCON BRAND URETHRAL CATHETER KIT (Female) nurse assist incorporated 25 ea/Box 800-649-6800 3400 Northern Cross Blvd. Fort Worth, TX 76137 www.nurseassist.com Tray Label: REORDER NO. 7402 STERILE UNLESS OPENED OR DAMAGED SINGLE USE WELCON BRAND Urethral Catheter Kit (FEMALE) nurse assist incorporated A6003, R1 800-649-6800 3400 Northern Cross Blvd. Fort Worth, TX 76137 www.nurseassist.com CONTENTS: Vinyl Gloves PVP Swabsticks Specimen Label Specimen Tube w/Cap 8 Fr Urethral Catheter Underpad Lube Jelly Intended use: The trays are urological catheter trays.
- Product code
- FCM Tray, Catheterization, Sterile Urethral, With Or Without Catheter (Kit)
- Reason for recall
- The trays contain Triad Group Sterile Lubricating Jelly packs recalled by Triad Group because of the possibility of inadequate sterilization.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- The firm, Nurse Assist Incorporated, sent a letter dated January 20, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory immediately; quarantine the product; identify their customers and notify them at once about the product recall (Nurse Assist will provide bright labels to affix to your affected inventory); provide their customers with labels as well, and complete and return the attached RECALL ACKNOWLEGEMENT form via fax to: 817-231-1500 or mail to: Nurse Assist, Inc., 3400 Northern Cross Blvd., Forth Worth, TX 76137. Only the product and lots identified on the attached list are affected. If you have any questions or to request labels, please call Nurse Assist Customer Service Monday through Friday, from 8:00 am to 5:00 pm Central Time at 1-800-649-6800.
- Quantity in commerce
- 2,750 cases
- Distribution
- Worldwide distribution: USA (nationwide) and countries including: Canada, Dubai and Saudi Arabia.
- Product codes and lots
- Lot Codes: 804267, 807144, 809218, 810176, 811080, 901176, 904078, 905025, 906126, 907194, 908132, 909104, 910154, 912285, 1002029, 1003201, 1004076, 1004264, 1005225, 1006299, 1008050, 1009374, and 1011140.