Z-2552-2011

IMPAX CV (Cardio Vascular) The IMPAX Cardiovascular suite is a cardiovascular information system, providing...

AGFA Healthcare Corp. / initiated 2011-03-21 / Terminated / root cause: manufacturing process

AGFA Corp. began a recall of IMPAX CV (Cardio Vascular) The IMPAX Cardiovascular suite is a cardiovascular information system, providing archiving, image display and modality/study... on . The reason FDA records is "Customer error with the angiography (XA: X-Ray Angiography) measurement calibration functionality within the Cardiovascular Review Station resulted in improper calibration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2552-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
IMPAX CV (Cardio Vascular) The IMPAX Cardiovascular suite is a cardiovascular information system, providing archiving, image display and modality/study specific structured reporting.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Customer error with the angiography (XA: X-Ray Angiography) measurement calibration functionality within the Cardiovascular Review Station resulted in improper calibration.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm, AGFA Healthcare, sent an "URGENT SAFETY NOTICE" letter dated March 21, 2011 via FED-EX to its customers.. The letter describes product, problem, and actions to be taken. The customers were instructed to distribute this information within their facility to all individuals who need to be aware and to complete and return the URGENT SAFETY NOTICE FEEDBACK FORM via fax to: 864-421-1664 or email. AGFA is supplying the customers with a CRS User Manual Addendum to emphasize the proper steps required to achieve an accurate measurement when using the calibration tool within CRS. Mitigation will provide the customer/end-user with awareness of the correct XA measurement calibration technique and the potential consequences if proper technique is not followed. Additionally, AGFA will offer customer training on the use of the calibration tool. This training will be provided remotely via the web at no cost. Customers were ask to register before May 1, 2011 by contacting AGFA Customer Support Center toll-free at 877-777-2432. Training sessions are from May-June 2011. If you have any questions about this matter, please contact your local AGFA HealthCare organization or the AGFA Customer Support Center at 877-777-2432.
Quantity in commerce
536
Distribution
Worldwide distribution: USA (nationwide) and country of Canada.
Product codes and lots
Software versions - 2.01.B10, 2.01.B15, 2.01.B25, 2.02.B32, 2.02.B39 2.02.B40, 2.02.B43, 2.03.B16, 2.02.B23, 2.03.B26, 2.03B31.P01, 2.03.B31, 2.04.B03, 2.04.B06, 2.05.B01, 2.04.B07, 2.06.47, 2.06.47.02, 2.05.B02, 2.06.47.08, 2.06.47.08, 2.06.47.08.03, 2.06.47.02, 2.06.47.18, 2.06.47.18.01, 2.06.47.27, 2.06.47.18.02, 2.06.47.30, 2.06.47.37, 2.06.47.40, 2.07.24, 2.08.05.01, 2.07.30.01, 2.07.30.02, 2.05.03.04, 2.09.03, 2.07.30.07, 2.07.35, 2.07.24.SLEH.07, 2.09.04.02, 2.07.35.01, 2.10.06, 2.11.06, 2.05.04.05, 2.12.09, 2.13.08, 2.13.08.SU1, 2.14.03, 2.14.03.SU1, 2.15.08, 2.14.03.SU2, 2.15.08.SU3, 2.16.08, 7.4.SU3, 7.8, 7.8.SU1, and 7.8.SU2