Z-0159-2015Class II

Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning...

Terumo Cardiovascular Systems Corporation / initiated 2011-03-01 / Terminated / root cause: design

Terumo Cardiovascular Systems Corp began a recall of Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog... on . FDA classified it as Class II and gives the reason as "Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0159-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog Codes: Kits with a 5-digit code beginning with the digit 7: Catalog Number: 7XXXX Product usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery. The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it.
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line.
Root cause tag
design
FDA root cause
Device Design
Action
Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.
Quantity in commerce
Catalog Number: 6XXXX: 4561 units; Catalog Number: 7xxxx : 19356 units
Distribution
Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.
Product codes and lots
Catalog Number: 6XXXX Lot Numbers: ML07, MM18, MM25, MN08, MN IS, MN22.  Manufacturing dates September 7, 20 I 0 through November 22, 20 I 0.  Catalog Number: 7XXXX Lot Numbers: ML07,ML20, MM18, MM25, MNO1, MN08, MN15, MN29,MP06, MP13. Manufacturing dates September 7, 2010 through December 13,2010.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0159-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0156-2015Class IITerminatedVoluntary: Firm initiated
Z-0157-2015Class IITerminatedVoluntary: Firm initiated
Z-0158-2015Class IITerminatedVoluntary: Firm initiated
Z-0159-2015Class IITerminatedVoluntary: Firm initiated
Z-0160-2015Class IITerminatedVoluntary: Firm initiated
Z-0161-2015Class IITerminatedVoluntary: Firm initiated