Z-2139-2011

***REF IOSS413*** OSSEIOTITE Certain Implant 4 x 13mm*** Sterile using Radiation*** Rx Only***BIOMET 3i...

ZimVie US Corp LLC / initiated 2011-03-10 / Terminated / root cause: manufacturing process

Biomet 3i, LLC began a recall of ***REF IOSS413*** OSSEIOTITE Certain Implant 4 x 13mm*** Sterile using Radiation*** Rx Only***BIOMET 3i Dental Iberica S.L.WTC Almeda Park, Ed, 1, Planta 1 Pl... on . The reason FDA records is "Biomet 3i is recalling their product Osseotite Certain Implant Device IOSS413. The internal thread was not manufactured correctly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2139-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
***REF IOSS413*** OSSEIOTITE Certain Implant 4 x 13mm*** Sterile using Radiation*** Rx Only***BIOMET 3i Dental Iberica S.L.WTC Almeda Park, Ed, 1, Planta 1 Pl. de la Pau s/n08940 Cornella de Llobregat (Barcelona) Spain*** Biomet 3i Dental Implants are intended for surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.
Product code
O-FY
Reason for recall
Biomet 3i is recalling their product Osseotite Certain Implant Device IOSS413. The internal thread was not manufactured correctly.
Root cause tag
manufacturing process
FDA root cause
Other
Action
BIOMET 3i sent an Urgent Medical Device Recall Letter on March 14, 2011, to all affected customers via emial or fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers were told to contact Customer Service at 1-800-342-5454. Customers were instructed to respond with the attached Response Fax or email [email protected]. If customers still had implant(s) in their inventory, they were to return the implant(s) to BIOMET 3i using the following address: BIOMET 3i Regulatory Compliance Department 4555 Riverside Drive Palm Beach Gardens, Florida 33410 Incident No. CD117886 A replacement implant(s) will be sent to them one the product is returned. For questions regarding this recall call 561-776-6906 (EST 8:00am to 5:00pm).
Quantity in commerce
217
Distribution
Worldwide Distribution - USA including AK, FL, GA, NC, NJ, NY, SC and the countries of Canada, Europe, Japan, and South America.
Product codes and lots
Lot number: 2009110838.