Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2104-2011

Biomet Interlok 75 MM fixed I-Beam Tibial Plate with locking bar CO-CR-MO/TI 6AL 4V ALLOY, REF 141224...

Zimmer GmbH / initiated 2011-03-03 / Terminated / root cause: packaging

Biomet, Inc. began a recall of Biomet Interlok 75 MM fixed I-Beam Tibial Plate with locking bar CO-CR-MO/TI 6AL 4V ALLOY, REF 141224, Sterile, BIOMET ORTHOPEDICS, WARSAW. IN I-Beam Tibia... on . The reason FDA records is "The firm is recalling the product due to the product's packaging being labeled "I-Beam 75mm. Tibial Plate", but the package actually contains the "Interlok 67mm. CR Tibial Plate"". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2104-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Interlok 75 MM fixed I-Beam Tibial Plate with locking bar CO-CR-MO/TI 6AL 4V ALLOY, REF 141224, Sterile, BIOMET ORTHOPEDICS, WARSAW. IN I-Beam Tibia Plates are knee joint replacement prostheses intended for application with or without bone cement.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The firm is recalling the product due to the product's packaging being labeled "I-Beam 75mm. Tibial Plate", but the package actually contains the "Interlok 67mm. CR Tibial Plate".
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Biomet Orthopedics Inc initiated an electronic communication with Direct Accounts on March 3, 2011 and instructed them to quarantine and return all remaining/unused devices to the firm. For questions regarding this recall call 574-372-3983.
Quantity in commerce
8
Distribution
Worldwide Distribution -USA including FL, IN, and NC and the countries of Japan and the Netherlands
Product codes and lots
REF 141224, LOT J100921070